THE APPLICATION OF SHIFT METHOD USING MOVING LASER ON CANCER CASES IN HEAD AREA USING 3DCRT AND VMAT TECHNOLOGY
ABSTRACT
The accuracy of patient position is essential to achieve the purpose of radiotherapy. Therefore, before radiotherapy, the isocenter of the
patient must be determined to find out the target location. Verification serves to determine the isocenter that shifts from the reference point
on the CT simulator. This research was conducted using field observation method, comparing the results of the isocenter after verification with the isocenter on the moving laser. Eight (8) samples from primary data were taken. Shift data was obtained from the average value and standard deviation. The instrument used was a
linear accelerator of clinic iX type and On-Board Imager. The results obtained were the average lateral coordinate point of 0.055 cm; at the longitudinal isocenter point of 0.10 cm; and at the vertical isocenter point of 0.053 cm. It was concluded that the average value o the isocenter shift was less than 0.3 cm. This method can be used because the shift does not exceed the tolerance limit.
Keywords: moving laser, verification, on-board imager
Nursama Heru Apriantoro - Personal Name
Samsun - Personal Name
Rudi Kurniawan - Personal Name
Agnes Dwika Septiti - Personal Name
Samsun - Personal Name
Rudi Kurniawan - Personal Name
Agnes Dwika Septiti - Personal Name
2021
001
001
ebook
Indonesia
Universitas Negeri Jakarta
2019
SPEKTRA: Jurnal Fisika dan Aplikasinya; p-ISSN: 2541-3384 ;e-ISSN: 2541-3392 ;Volume 4, Issue 3, December 2019 ;url : doi.org/10.21009/SPEKTRA
Author : Samsun-1, Nursama Heru Apriantoro-2, Rudi Kurniawan-3, Agnes Dwika Septiti-4
Department of Radiodiagnostic and Radiotherapy Engineering, Health Polytechnic of Jakarta II,
Jl. Hang Jebat III Block F3 Jakarta Selatan 12720; kd17-09-2021ns
347
LOADING LIST...
LOADING LIST...
Nursama Heru Apriantoro. (2019).THE APPLICATION OF SHIFT METHOD USING MOVING LASER ON CANCER CASES IN HEAD AREA USING 3DCRT AND VMAT TECHNOLOGY.(Electronic Thesis or Dissertation). Retrieved from https://localhost/setiadi